Vicat Dry

Vicat Dry

Vicat Dry

Product catalog summary
Overview
The document provides technical information about the Vicat Dry instrument, used for determining the Vicat softening temperature (VST) of thermoplastics as per ISO 306. The instrument offers an oil-free, automatic testing process, ensuring clean and odor-free conditions.

Applications
The Vicat Dry is used in research and development, goods inwards checks, and production monitoring.

Advantages and Features
  • Easy and convenient operation.
  • Automatic test sequence with different versions available (two or six testing stations).
  • Oil-free measuring principle.
  • Test-sequence parameterization and measured-value display via PC.
  • Excellent agreement with conventional oil-bath instruments.

Function Description
The instrument automatically applies the test weight, heats the specimens, and measures penetration. It features automatic re-cooling and specimen removal.

Functions, Elements, and Interfaces
  • Microprocessor-controlled electronics for temperature control.
  • Automatic compressed-air re-cooling and fluid cooling connection.
  • Safety features including a thermostat and switch.
  • PC connection via RS232 interface for test control and data acquisition.
  • Integrated weight sets of 10N and 50N.

Technical Data
  • Display accuracy at 250 °C: ±0.5 °C.
  • Temperature measurement resolution: 0.1 °C.
  • Heating rate: 50, 120, or freely adjustable K/h.
  • Displacement measurement range: approx. 2.5 mm.
  • Re-cooling times: 15 min with water, 25 min with air.
  • Power requirements: 230/240 V, 50/60 Hz, 550 W.

Specimen Data
Maximum VST specimen dimensions are 10 x 6.5 x 10 mm.

Optional Accessories
  • Specimen cutters for dimensioning Vicat specimens.

Note
Technical changes may occur during ongoing development. All data is provided at ambient temperature.
See more

Catalog excerpts

Vicat Dry-1

Materials testing Product Information Vicat Dry Vicat specimens are simply placed on the measuring station Applications Determination of Vicat softening temperature (VST) • of thermoplastics as per ISO 306 On completion of preheating under the influence of force, the displacement signals are zeroed (programcontrolled) and warming of the heating block at the preset rate is started. As soon as the required penetra‐ tion is attained at all stations, the heating is switched off and the test is ended. No more irritating oil fumes! Advantages and features Vicat Dry instruments feature easy, convenient opera‐ tion. Particular areas of use include research and devel‐ opment, goods inwards checks and production moni‐ toring. The oil-free measuring principle ensures clean, odor-free working conditions. The test sequence is completely automatic. Different versions of the instrument are available, with two or six testing stations according to requirements. Testsequence parameterization and measured-value display can be performed conveniently on a connected PC. The results of the Vicat Dry test display excellent agree‐ ment with those obtained using a conventional oil-bath instrument. The same is true of test-result consistency. Automatic test sequence Function description Once the heating block has reached the starting temperature, the specimens are positioned in the meas‐ uring stations, the instrument is closed pneumatically and the test weight specified in the standards is auto‐ matically applied. Re-cooling is started automatically; the specimens can afterwards be removed. Functions, elements and interfaces • built-in microprocessor-controlled electronics for temperature control and measured-value acquisition • fully automatic test-sequence • automatically controlled compressed-air re-cooling • connection for fluid cooling • safety thermostat • low heating of housing • safety switch prevents crush risk • serial interface for connecting a PC for test-control and data acquisition via Zwick Roell testing software • integrated compensation for heat expansion • integrated weight sets, 10N and 50N Re-cooling the heat-transfer liquid The instruments are equipped with compressed-air cooling as standard. If sufficiently cool compressed air is not available, the inst

 Open the catalog to page 1
Vicat Dry-2

Zwick Materials testing Zw/ck / RoellProduct Information Vicat DryTechnical data Description Value General technical data for Vicat Dry Display accuracy at 250 °C ±0.5 °C Spatial temperature gradient ±0.2 °C Temperature measurement resolution 0.1 °C Temperature sensor 2x measurement of heating-block temperature, 1 additional specimen-adjacent sensor per measuring station Heating rate 50, 120 or freely adjustable K/h Displacement measurement with inductive LVDT measuring transducers Displacement measurement range, approx. 2.5 mm Re-cooling times (dependent on...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.