NanoVACQ Pressure and Temperature - 1

Data sheet TMI-Orion NanoVACQ Pressure and Temperature Simultaneous pressure and temperature measurement. NanoVACQ Pressure and Temperature is a data logger equipped with 1 pressure sensor and up to 2 temperature sensors on the same logger, answering the needs of many industrial processes. Various NanoVACQ Pressure and Temperature models are described below and can vary by probe shape and length, operating range and capacity of the battery pack. Metrology Pressure operating range Temperature operating range From -55°C to +85°C From 30 mbar to 5 bar, 15 bar or 30 bar from -55°C to 140°C Possibility of higher pressure Temperature calibration uncertainty* Pressure calibration uncertainty* • • Temperature ± 0.1°C from -55°C to +140°C • Pressure 0.8 mbar (5 bar) (± 0.05°C upon 2.6 mbar request) (15 bar) • Temperature 0.008°C ± 10 mbar from 0°C to 140°C and from 30 mbar to 5 bar (1Hz or 10Hz mode) ± 12 mbar from 0°C to 140°C and from 30 mbar to 15 bar (1Hz mode) ± 17 mbar from 0°C to 140°C and from 30 mbar to 15 bar (10Hz mode) Unspecified from -30°C to 0°C Not functional from -55°C to -30°C *The specified uncertainties correspond to two standard deviations. The uncertainties are calculated taking into account the various significant error sources, including the calibration probes, the equipment, the environmental conditions, the influence of the logger, repeatability, etc… TMI-Orion NanoVACQ Pressure and Temperature Each logger can be calibrated and adjusted at the temperature poin

Open the catalog to page 1
NanoVACQ Pressure and Temperature - 2

Start set up: immediate or delayed Time stamped measurement data Memory set up: stop at maximum capacity or loop writing Battery level alert with Qlever software Technical specifications Model Internal Number of external temperature probe* channels Pressure probe type External temperature probe type Temperature probe dimensions Water ATEX tightness compliant NanoVACQ PT-Tc 1 rigid tip at the end of 1 flexible deport (Viton®) NanoVACQ PT-Td 1 rigid tip at the end of 1 flexible deport (Teflon®) * Internal platinum temperature sensor for pressure sensor compensation. TMI-Orion NanoVACQ...

Open the catalog to page 2
NanoVACQ Pressure and Temperature - 3

Technical specifications Material Logger body: 316L Stainless steel With 014Z battery pack With Routine HE battery pack With 014ZFL battery pack Temperature sensor Memory capacity 48 000 acquisitions divided by number of measurement channels Memory capacity with BigMemory 294 500 acquisitions divided by number of measurement channels 1 Hz Programmable: minimum 1 second, maximum 59 minutes and 59 seconds Acquisition rate Programmable: minimum 100 ms, maximum 59 minutes and 59 seconds Program duration Programmable: days, hours, minutes Programmable start: by date, hour, minute or on...

Open the catalog to page 3
NanoVACQ Pressure and Temperature - 4

Autonomy The NanoVACQ Pressure and Temperature is powered by a battery pack; its autonomy depends on environment and operational conditions of the application (extreme temperatures, data acquisition rate). As a result of the variety of environments and operational conditions, TMI-Orion does not guaranty the battery lifetime and recommends that the user determine the battery lifetime according to his own process conditions and experience. Software and related products NanoVACQ Pressure and Temperature is used with Qlever software. 7/8/10. Data transmission and visualization are done after...