Overview
The TMI-Orion High-T-Dry is a temperature and shrinkage data logger designed for monitoring drying processes in bricks, tiles, and ceramics, capable of withstanding temperatures up to 250°C.
Metrology
The device consists of a logger and a sensor plate connected by a high-temperature cable. The logger, which houses the electronics and a replaceable battery, can operate up to 125°C and is equipped with watertight connectors. The sensor plate, capable of withstanding up to 250°C, includes a temperature sensor and a retractometer.
Specifications
  • Logger Operating Range: 0°C to +125°C
  • Sensor Set Operating Range: 0°C to +250°C
  • Measurement Range: 0°C to +250°C
  • Resolution: 0.05°C
  • Temperature Uncertainty: ±0.5°C from 0°C to 140°C, ±0.7°C from 0°C to 250°C
  • Shrinkage Measurement: 20 mm movement, 140 mm to 160 mm range, 0.02 mm resolution, ±0.5 mm uncertainty
Technical Specifications
  • Material: 316 L Stainless Steel
  • Dimensions: Sensor plate - Length 169 mm, Width 55 mm, Height 52 mm; Logger body - Diameter 31 mm, Length 39 mm
  • Temperature Sensor: PT100
  • Shrinkage Measurement: Linear potentiometer
  • Memory Capacity: 48,000 acquisitions
  • Acquisition Rate: Programmable from 1 second to 59 minutes and 59 seconds
  • Power: User-replaceable battery pack
  • Connectivity: USB wired interface
Functions
  • Immediate or delayed start setup
  • Memory setup options
  • Time-stamped data
  • Battery level alert via Qlever software
Deliverables
  • High-T-Dry logger with battery pack
  • Calibration certificate
  • Configuration and calibration file
  • Optional USB interface, Qlever software, and transport case
Services
Annual preventative maintenance and calibration are recommended. User-replaceable battery packs are available.
Autonomy
The battery life depends on environmental and operational conditions, and users are advised to determine battery life based on their specific processes.
Software
The High-T-Dry operates with the Qlever software platform for data acquisition, management, and visualization, compatible with Windows Vista/7/8/10.
See more

Catalog excerpts

High-T-Dry-1

DATA SHEET TMI-Orion High-T-Dry Temperature and shrinkage data logger The High-T-Dry is an autonomous data logger equipped with a temperature sensor and a rectractometer. It is made in two parts, the logger and the sensor plate, linked together by a high-temperature cable: • The sensor plate contains a temperature sensor and a retractometer fixed to a metal plate which can withstand temperatures up to +250°C. The High-T-Dry correlates the air temperature changes with the shrinkage of bricks during industrial drying. It also enables the acute evaluation of drying within two parts of the same brick or tile. Each logger can be calibrated and adjusted at the temperature points corresponding to the user’s needs. (*) The specified uncertainties correspond to two standard deviations. The uncertainties are calculated taking into account the various significant error sources, including the calibration probes, the equipment, the environmental conditions, the influence of the logger, repeatability, etc... TMI-Orion High-T-Dry The logger contains the electronics and a user-replaceable battery pack. It can be exposed to temperatures not exceeding +125°C. It is equipped with watertight connectors and can be immersed in a water thermal shield.

 Open the catalog to page 1
High-T-Dry-2

• Start set up: immediate or delayed • Time stamped data • Memory set up: stop at maximum capacity or loop • Battery level alert with Qlever software TMI-Orion High-T-Dry

 Open the catalog to page 2
High-T-Dry-3

Autonomy The High-T-Dry is powered by a battery pack; its autonomy depends on the environment and the operational conditions of the application (extreme temperatures, acquisition rate). As a result of the variety of environments and operational conditions, TMI-Orion does not guaranty the battery lifetime and recommends that the user determine the battery lifetime according to his own process conditions and experience. Software and related products The High-T-Dry operates with Qlever software platform. Qlever software platform: data acquisition, management and visualization of data from TMI-Orion...

 Open the catalog to page 3

Archived catalogs

  1. NanoVACD AD

    3  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.