CPA2501 - 1

Calibration technology Air data test indicator Model CPA2501 WIKA data sheet CT 29.02 Applications ■■ Aviation calibration laboratories ■■ Aviation repair stations ■■ Manufacturers of avionics equipment ■■ Manufacturers of aerospace equipment ■■ Wind tunnels Special features ■■ Accuracy up to 0.01 % FS ■■ RVSM compliant ■■ Ps, Qc, Ps/Pt or Ps/Qc configuration with virtual channels ■■ Altitude and airspeed rate indication Air data test indicator, model CPA2501 ■■ Altitude ranges to 100,000 ft. and airspeed ranges to 1,150 Description Applications The model CPA2501 air data test indicator is used in aviation and aerospace applications for calibration of altimeters and airspeed indicators, displaying altitude, airspeed, altitude rate (vertical airspeed) and airspeed rate (acceleration). Applied wherever there is a need for a high level of accuracy in an avionics indicator or calibration instrument. Functionality The CPA2501 can be configured with an altitude channel and an airspeed channel consisting of Ps/Pt or Ps/Qc. It can be configured as a single-channel altitude/altitude rate indicator, (Ps only) a single-channel airspeed/airspeed rate indicator (Qc only) or it can be a dual-channel unit configured as Ps/Pt or Ps/Qc. In the Ps/Pt version the airspeed/airspeed rate channel is a calculated channel. An optional barometer can be installed for display in the front panel or as a separate channel. Pressure ranges for each channel can be specified by the customer. WIKA data sheet CT 29.02 ∙ PN 0019643001A ∙ 01/2017 Data sheets showing similar products and accessories: Precision pressure sensor, models CPT6100, CPT6180, see data sheet CT 25.10 Precision pressure indicator, model CPG2500, see data sheet CT 25.02 Versions The CPA2501 versions are as follows: ■■ Ps (altitude version) ■■ Qc (airspeed version) ■■ Ps/Qc with a virtual Pt channel ■■ Ps/Pt with a virtual Qc channel An optional barometric sensor can be added to any version. Communication The manual user interface is through a wide screen display with touchscreen. Navigation within the intuitive menu structure is easily learned. Recognizable icons, when touched, open screens for configuration and calibration. Communicating to a remote computer is achieved through Ethernet, RS-232 or optional IEEE-488 communications. Commands used for communication are the same as the previous version of Mensor avionics gauge models

Open the catalog to page 1
CPA2501 - 2

WIKA data sheet CT 29.02 • PN 0019643001A • 01/20

Open the catalog to page 2
CPA2501 - 3

Base instrument 3) Calibration in a horizontal position. Approvals and certificates, see website WIKA data sheet CT 29.02 • PN 0019643001A • 01/2017

Open the catalog to page 3
CPA2501 - 4

1 Connection for optional barometric reference 2 Port Ps 3 Port Qc 4 Port Pt / Qc 5 Power supply 6 USB interface (host) 7 Ethernet port 8 USB interface (instrument) 9 RS-232 interface 10 IEEE interface 11 External sensor connection WIKA data sheet CT 29.02 • PN 0019643001A • 01/2017

Open the catalog to page 4
CPA2501 - 5

Dimensions in mm (in) Desktop case Front view Side view (left) 19" rack mounting, front view Single instrument Dual instrument WIKA data sheet CT 29.02 ∙ PN 0019643001A ∙ 01/2017

Open the catalog to page 5
CPA2501 - 6

Operator interface :! Menu ▼ Reading for optional barometer Local operation The intuitive operator interface of the CPA2501 provides visibility of one, two or three channels, each with or without the auxiliary display. Readings from the optional barometer can also be displayed in the lower right-hand corner. Pressure units for each channel and the barometer can be selected from a list of 38 metric and imperial units. The setup “apps” are continuously visible for fast configuration of various applications. Remote operation Remote control of the CPA2501 is achieved through the use of the...

Open the catalog to page 6
CPA2501 - 7

Reference pressure sensors Interchangeable sensor CPR8001 One or two pressure sensors can be chosen (see specifications). The interchangeable sensor CPR8001 are a special feature of the air data test indicator. Interchangeable sensors guarantee long-term operation with practically no downtime. The sensor can be removed for calibration and can be replaced by a freshly calibrated sensor. The ability to remove a sensor for calibration and replace it with a freshly calibrated sensor while the instruments remains in service saves time and money. Simply remove the four slotted screws on the rear...

Open the catalog to page 7
CPA2501 - 8

■■ Air data test indicator, model CPA2501 ■■ Reference pressure sensor model CPR8001 ■■ 19" rack mounting with side panels for a single instrument ■■ DKD/DAkkS calibration certificate ■■ 19" rack mounting with side panels for two instruments side ■■ Robust transport case ■■ Pressure adapters Ordering information Model / Case type / Reference pressure sensor / Barometric reference / Type of certificate for barometric reference / Additional ordering information © 2015 WIKA Alexander Wiegand SE & Co. KG, all rights reserved. The specifications given in this document represent the state of...

Open the catalog to page 8

All Mensor catalogs and technical brochures

  1. WIKA Handbook

    436 Pages

  2. Calibration

    20 Pages

  3. Custom Systems

    20 Pages

  4. Model CPB5000

    12 Pages

  5. Model CPB3800

    15 Pages

  6. Model CPG500

    6 Pages

  7. Model CPG1500

    11 Pages

  8. CPG2300

    2 Pages

  9. CPA8001

    8 Pages

  10. CPC8000

    8 Pages