Overview: The Models 6500 & 6500-HA Automated Air Data Test Sets are designed for testing and calibrating various aircraft instruments such as altimeters, airspeed/Mach indicators, and flight data recorders. They require a pressure and vacuum source and are noted for their simplicity and high accuracy, meeting RVSM requirements with annual calibration.
Specifications:- Static Output: The 6500 model offers a pressure range of 0.1 to 42 inHg with a resolution of 0.001 inHg and accuracy of 0.002 inHg. The 6500-HA model provides a range of 0.1 to 38 inHg with a resolution of 0.0001 inHg and accuracy of 0.001 inHg. Altitude and climb parameters are also specified with varying resolutions and accuracies.
- Pitot Output: Both models support a pressure range of 0.1 to 100 inHg. The 6500-HA offers higher resolution and accuracy. Airspeed and Mach parameters are detailed with specific ranges and accuracies.
- EPR Function: Range from 0 to 199 with a resolution of 0.001 and typical accuracy of 0.002.
Features:- Jog Feature: Allows incremental adjustments using arrow keys.
- Profiles Feature: Enables downloading and storing up to 20 profiles for simplified operation.
Requirements:- Vacuum: Requires a vacuum pump capable of 0.3 inHg absolute or better.
- Pressure: Requires dry air at 40 psig with a dedicated pressure regulator.
- Power: Operates on 90-260 VAC, 48-440 Hz, 100 VA.
Interfaces and Design: Includes RS232 and IEEE-488 interfaces, with optional encoder and remote unit. The design is a 4U rack-mount, measuring 7.0" high x 19.0" wide x 20.0" deep, weighing 25 lbs.
Environmental Specifications: Operating temperature ranges from 0°C to 50°C, storage temperature from -25°C to 75°C, and humidity from 5 to 95% non-condensing.
Contact Information: Laversab, Sugar Land, TX, USA. Phone: (800) 285-7337, (281) 325-8300. Fax: (281) 325-8399. Website: www.laversab.com. Email:
[email protected].