DCN1000N/H series

DCN1000N/H series

DCN1000N/H series

Product catalog summary
Overview
The DCN1000N/H series are advanced low-temperature infrared reference sources designed for precise temperature regulation and stability. They are ideal for validating and characterizing infrared sensors, including high-resolution thermographic cameras and long-range thermal imagers.

Specifications
- Temperature Range: Absolute mode from -15°C to +150°C (extendable to +200°C); Differential mode from -35°C to +130°C.
- Emissive Area: Ranges from 50 mm x 50 mm to 300 mm x 300 mm.
- Thermal Uniformity: ±0.01°C at ambient ±5°C.
- Emissivity: 0.98 ± 0.02, with an option for 0.99 ± 0.01.
- Stability: 0.5 mK.
- Temperature Measurement Accuracy: ±0.01°C in differential mode, ±0.03°C in absolute mode.
- Display Resolution: 0.0001°C.
- Remote Control: Via Ethernet, RS232, and IEEE488 interfaces.

Features
- High stability and thermal uniformity.
- Real-time temperature data display with an intuitive interface.
- Silent operation and compact emissive head.
- Radiometric calibration over multiple bandwidths.
- Remote control capabilities and built-in test equipment.

Options and Accessories
- Motorized target wheel and multiple accessories.
- Enhanced accuracy in absolute mode (+/-0.01°C).
- Climatic chamber operating conditions.
- Enhanced emissivity (>0.99).
- LabVIEW drivers for communication interfaces.
- Anti-condensation and frost system.

Applications
The DCN1000N/H series is suitable for a wide range of automated tests, including NETD, temporal noise, fixed pattern noise, MTF, FoV, distortion, spatial resolution, MRTD, and TOD.

Contact Information
- Headquarters: HGH SYSTEMES INFRAROUGES, Igny, France.
- US Office: ELECTRO OPTICAL INDUSTRIES, Goleta, CA, USA.
- Asia Office: ASIA INFRARED SYSTEMS, Singapore.
See more

Catalog excerpts

DCN1000N/H series-1

> Infrared validation test equipment DCN1000N/H series ABSOLUTE AND DIFFERENTIAL TEMPERATURE THE MOST ACCURATE LOW TEMPERATURE BLACKBODIES The DCN1000N/H extended area blackbodies are low temperature infrared reference sources operating either in absolute or differential mode. Featuring the highest available stability of regulation, they are particularly well adapted for the characterization and performance validation of a very wide range of IR sensors, such as high resolution cameras for thermography, and long range thermal imagers. They consist of an emissive head of various sizes whose temperature is accurately controlled via a 2U electronic unit through an ergonomic interface. The temperature of the high emissive surface is stabilised within 0.5mk at temperatures above and below ambient temperature. Just as any other HGH blackbody, the DCN1000N/H family are provided with a radiometric certificate of calibration demonstrating the reliability of this IR reference source for two years. Targets can be added, as well as the INFRATEST software, thanks to which a wide range of tests can be automated: NETD, temporal noise, fixed pattern noise, MTF, FoV, distortion, spatial resolution, MRTD, TOD, etc. • Extended areas up to 300 mm x 300 mm • Differential and absolute modes operation • Real time display of temperature data • Intuitive interface • The highest stability: < 0.5mk • High thermal uniformity and emissivity • Silent operation • Compact emissive head • Absolute temperature range from –15 °C to +150 °C • Control through coloured touchscreen panel • Radiometric calibration over multiple bandwidths • Remote control via Ethernet link, RS232, IEEE488, WiFi • Built-in test equipment (BITE) • Infratest - LT control software > DCN1000 & motorised target wheel OPTIONS • Absolute temperature range up to +200°c • Motorised target wheel and multiple accessories • Targets for NETD, LSF/MTF, MRTD, distortion, etc. • NETD, LSF/MTF, MRTD and TOD calculation software • Enhanced accuracy of absolute mode (+/-0.01°C) • Climatic chamber operating conditions • Enhanced emissivity (>0.99) OPTIONS • LabVIEW drivers for all communication interfaces • Anti-co

 Open the catalog to page 1
DCN1000N/H series-2

DCN1000N/H series ABSOLUTE AND DIFFERENTIAL TEMPERATURE MBLACKBODIES Infrared validation test equipment DCN1000 - coffimator reference source US Office ELECTRO OPTICAL INDUSTRIES 320 Storke Rd., Ste. 100 Goleta, CA 93117, USA Phone: 805.964.6701 Fax: 805.967.8590 Email: [email protected] Asia Office ASIA INFRARED SYSTEMS 541 Orchard Rd., #09-01 Liat Towers Singapore 238881 Phone: +65 6933 1394 Email: [email protected]

 Open the catalog to page 2

All HGH Infrared Systems catalogs and technical brochures

  1. ISV-series

    2  Pages

  2. TwiN1000

    2  Pages

  3. RCN

    2  Pages

  4. DCN1000W-L

    2  Pages

  5. ECN100 series

    2  Pages

  6. Spynel-M

    2  Pages

  7. Spynel-U

    2  Pages

  8. Spynel-C

    2  Pages

  9. Spynel Series

    2  Pages

  10. Spynel-X

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.