Overview: The PACE1000 is a high precision Druck pressure indicator suitable for various applications such as test bench, bench top, and panel mount calibration, test, and monitoring. It utilizes advanced piezo-resistive and resonant pressure sensor technology.
Key Features:
Options and Applications:
Specifications:
Ordering Information: Customers should specify model, options, area of use, and precision levels when ordering. Various internal sensor configurations and optional features are available.
Product Specifications:
Supporting Services:
Product Range:
Related Products: Druck offers a wide range of pressure transducers, transmitters, indicators, calibrators, controllers, and Air Data test systems. Portable calibrators also cover temperature and electrical parameters.
For more information, visit druck.com.
PACE1000Precision pressure indicatorA high precision Druck pressure indicator, designed for test bench, bench top and panel mount calibration, test and monitoring applications.Features • Intuitive icon driven task menu structure • Switch Test/Analogue Output option • Switch Test/Voltage Free Contact option • Aeronautical option • Leak Test option • RS232, IEEE connectivity, Ethernet and USB as standard • Min/Max/Average display • Compatible with software packages • 28 selectable pressure units plus 4 user defined units • Various service support options available • Single or multiple pressure range configurations • Up to three pressures displayed simultaneously • Utilises Druck's new unique range of piezo-resistive and Resonant Pressure Sensor technology • Full scale pressure ranges up to 1000 bar (14500 psi/100 MPa) • Long term stability up to 0.001% FS • Barometric reference option • Airfield Task as standard with barometric option: Display QFE, QFF or QNH in pressure units or as altitude in feet or meters • Negative gauge calibration included as standard • Data Logging as standard with on screen replay • Selectable numeric or graphic display • High resolution touch
Open the catalog to page 1PACE precision pressure indicator The new PACE precision pressure indicator brings together the latest measurement technology from Druck to offer an elegant, flexible and economical solution to pressure measurement for test, calibration and monitoring. PACE employs digitally characterized pressure sensors which offer the quality, stability and precision associated with this latest generation of piezo-resistive and resonant devices. PACE1000 offers a wide choice of standard pressure ranges, with three levels of precision to ensure that PACE1000 can accommodate specification and budget requirements....
Open the catalog to page 2Specifications Pressure measurement IPS pressure ranges: See section 5 on page 6 for full list of pressure ranges available 0.01% FS includes linearity, hysteresis, repeatability and temperature effects. For down ranged variants: 1300 mbar Precision = IRS1 high precision 0.005% FS includes linearity, hysteresis, repeatability and temperature effects. For down ranged variants: 1300 mbar Precision = 0.0077% FS; 2600 mbar Precision = 0.0067% FS 0.0025% FS includes linearity, hysteresis, repeatability and temperature effects. For down ranged variants: 1300 mbar Precision = IRS2 premium precision...
Open the catalog to page 3Absolute ranges 2000, 3500 mbar accuracy (2 Sigma) over calibrated temperature range 0.0004% RDG + 0.0027% FS. Includes measurement precision, measurement long term stability (see below) and calibration equipment expanded uncertainty. Pseudo gauge range accuracy (3.5bara and below) will need to include the barometer uncertainty using the RSS (root sum of squares) method. Pseudo absolute mode precision Gauge mode precision + barometric precision Pseudo gauge mode precision absolute mode reference precision + barometric precision *To meet annual IRS3 accuracy specification, zeroing against a barometric...
Open the catalog to page 4Ordering information Please state the following (where applicable) Please state area of use for mains lead: PACE chassis *Please state precision for each IPS/IRS selected. • • IRS3-B must be included on ranges >=8bara to maintain datasheet specification therefore PACE1002/1003 should be selected for these ranges South Africa/India China The range of optional features includes: • Leak Test - Automatically measures leak rates in the desired units/minute or units/seconds • Switch Test/Analogue Output - Accurate calibration of pressure switches/integration into older ATE applications • Switch Test/Voltage...
Open the catalog to page 5+211 bara 3,060 psia 21.1 MPaa
Open the catalog to page 6PACE barometric option*7. Supporting services PACE Internal Resonant Sensor Barometer ranges - IRS-B Services ordering information: Please order the following as separate line items: Part number Description External IDOS universal pressure module 6. Physical accessories Please order the following as separate line items: Part number PACE1000 - Precision Pressure Indicator PACE1001 Barometer - Precision Barometric Indicator and Recorder PACE5000 - Single Channel Pressure Controller Chassis PACE6000 - Dual Channel Pressure Controller Chassis CM0 - Standard precision high speed pressure controller...
Open the catalog to page 7Delivering world class pressure measurement and calibration technology 2026 Copyright by Druck Limited. All rights reserved.
Open the catalog to page 86 Pages
17 Pages
7 Pages
6 Pages
8 Pages