RPS/DPS 8000High accuracy resonant pressure sensor Since 1972, Druck has manufactured precision pressure sensors with a capability to meet critical applications in industrial, aerospace, oil and gas, and research environments. Today, Druck is part of Baker Hughes and has continually worked to develop and improve on the performance of our pressure sensors to meet our customers' requirements. The RPS/DPS 8000 is the first product to incorporate the TERPS technology over 10 years ago. TERPS is a resonant silicon pressure sensor technology platform that provides an order of magnitude greater accuracy and stability than current pressure measurement technologies. The TERPS technology also extends the pressure range capability to high pressures and, by incorporating true pressure media isolation, greatly improves its suitability for use in harsh environments. In addition to providing the performance and packaging improvements available with TERPS, the RPS/DPS 8000 product line takes advantage of best practices to offer a wide range of pressure and electrical connections to enable a level of customization for your specific requirements never before available in the performance class of this sensor. The combination of the power of the TERPS technology and the quality, reliability and flexibility of the RPS/DPS 8000 Series offers a truly unique solution for high accuracy and high stability pressure measurement requirements. • High Precision, ±0.01% FS over compensated temperature range • High Stability, ±100 ppm FS/year • Wide temperature range, up to -55°C to +125°C (-40° to 257°F) • Media isolated construction, suitable for use in harsh environments • Multiple output configurations, RS-232, RS-485, USB 2.0, CANbus, frequency and diode (TTL) • Wide selection of pressure and electrical connections to suit specific requirements
Open the catalog to page 1Any zero-based range between 1 and 200 bar (14.5 to 2900 psi) can be specified. (Performance will be of the full scale of the base pressure range selected.) Barometric ranges are available in the RPS/DPS 8100 series. The lowest calibrated pressure is 35 mbar absolute. Sensor failure pressure Pressure containment • For frequency and diode output, the above accuracies are achievable by using a polynomial curve fit algorithm and coefficient data supplied with sensor. • Sensors are calibrated against standards traceable to UKAS operating to better than 100 ppm. Compensated temperature ranges There...
Open the catalog to page 2All temperature effects are included in the accuracy statement. Note: Unless otherwise specified, specifications are at the reference conditions of 25°C (77°f) ±5°C (±9°F) and 1 bar (14 psi). Less than 0.2 mbar/g Storage temperature range As compensated temperature range. As compensated temperature range. Pressure media Fluids compatible with 316L stainless steel and Hastelloy C276 Not suitable for media that has an oxygen concentration >21% or other strong oxidizing agents. This product contains materials or fluids that may degrade or combust in the presence of strong oxidizing agents. Ingress...
Open the catalog to page 3Connection details Electrical
Open the catalog to page 4a. MALE PRESSURE CONNECTION MEDIUM PRESSURE CONSTRUCTION Core construction ranges greater than 104 bar except electronics options F, G & V DPS8000 core construction (electronics options F, G & V) 1. All dimensions are nominal lengths and are subject to change. 2. All dimensions are in millimeters (inches). 3. Other pressure and electrical connectors may be available, please contact Druck.
Open the catalog to page 5Ordering information1. Select model codeMain product variant RPS Resonant pressure sensor - frequency and diode output (Note 1) DPS Digital pressure sensor - digital output (Note 1) Diameter, material and isolation 80 25 mm stainless steel oil isoliated Electrical connect 0 No electrical connector (flying leads) (Note 6) 3 Polyurethane cable (depth) IP68 4 Hytrel cable (depth) IP68 C 1/2" NPT conduit with polyurethane cable (non-exd only) H Orange PTFE cable U USB-C socket (Note 3) Electronics option (Note 8) Compensated temperature range TA -10 to +50 °C Full thermal calibration Hazardous area...
Open the catalog to page 6Ordering information (cont.) Note 1: RPS variants require electronics option code ‘1’. DPS variants require electronics option code ‘F’, ‘G’, ‘C’ or ‘V’. Note 2: Text goes here Note 2: Requires electronics option code ‘1’, "F" or ‘C’ and Electrical Connector Codes ‘0’, ’2’, ’6’, ’G’, or ’H’. Note 3: Only available with USB electronics option. Note 4: Only available with USB-C socket connector. Note 5: Choose one of these connectors for pressure ranges above 104 bar. Note 6: Flying lead length >200 mm Note 7: Available on ranges > 2 bar only. Note 8: Maximum cable length: (1) Frequency & Diode...